|
The performance of a process instrument such as a temperature or pressure sensor normally depends on its accuracy and response time. Accuracy is a qualitative term that describes how well the instrument may measure the process parameter (see Section 1.4 for an in-depth discussion), and response time specifies the speed by which the instrument can detect any significant change in the value of a process parameter. Accuracy and response time are largely independent and are therefore identified through separate procedures. The accuracy of a process instrument is established through its calibration, and the response time is determined by exposing the instrument to a dynamic input and measuring its response time from the transient output. Response time is discussed in more detail in Section 1.9. Calibration is done by providing the instrument with a number of known and stable inputs to ensure that the output accurately represents the input. Two terms are important in instrument calibration. These terms are zero and span , as illustrated in Figure 1.8a. In this figure, the calibration of a linear transmitter is approximated with a straight line, represented by the equation y = mx + b , where y is the output, x is the input, m is the slope of the line, and b is the intercept. The calibration of an instrument may change due to a change in zero, a change in span, or a change in both zero and span. A change in zero is also referred to as a bias error, DC offset , or zero shift. A zero shift results in a change in instrument reading (either positive or negative) at all points along its range (Figure 1.8b). A zero shift can result from several causes, such as a change in ambient temperature affecting the calibration. For example, if an instrument is calibrated at room temperature and used at a different temperature, its output may include a bias error (or zero shift) due to the temperature difference. The change in span is also referred to as a gain error or span shift. A span shift means an increase or a decrease in the slope of the instrument output line for the same input (see Figure 1.8c). Typically, calibration errors involving span shift alone are less common than calibration errors due to both zero and span shifts. In Figure 1.8c, both cases are shown: span shift without zero shift and span shift with zero shift. In pressure transmitters, about 40% of the calibration changes are caused by zero shift, about 30% by span shift, and only about 20% by span shift alone. 1 As for the remaining 10%, the calibration changes are due to other effects, such as nonlinearity.
Calibration of Pressure Sensors Calibration of pressure sensors (including both absolute and differential-pressure sensors) involves using a constant pressure source such as a deadweight tester (see Figure 1.8d). With a deadweight tester, constant pressure is produced for the sensor while the sensor output is monitored and adjusted
a pressure that corresponds to 100% of the span is applied by the deadweight tester, and the sensor output is adjusted to 20 mA. These adjustments to the output are made by setting two potentiometers provided in the pressure sensor. These adjustment devices are referred to as the zero and span potentiometers. The next step in the calibration of a pressure transmitter is to apply known pressures between 0 and 100% of span to verify the linearity of the transmitter and to make any necessary adjustments to obtain accurate mA outputs for all inputs. The zero and span adjustments of a pressure sensor interact, meaning that changing one will cause the other to change, and vice versa. Thus, in calibrating a pressure sensor, the zero and span are often both adjusted to produce the most accurate output that can be achieved for each input pressure. Because of the nonlinearities in some pressure sensors, the input/output relationships cannot be exactly matched, no matter how well the span and zero adjustments are tuned together. For that reason, in most pressure sensors, a linearity adjustment is also provided (in addition to the zero and span potentiometers) to help achieve the best agreement between the input pressure and the output current. In lieu of a deadweight tester, one can also use a stable pressure source and a precision pressure gauge as the input. Precision pressure gauges are available in a variety of ranges from a number of manufacturers (see Section 5.11). Highly accurate digital pressure indicators can also be used for calibration. As will be seen later, automated pressure sensor calibration equipment is also available that uses digital technology to offer both accuracy and convenience. As-Found and As-Left Data The calibration of an instrument can change with time. Therefore, instruments are recalibrated periodically. The periodic calibration procedure typically involves two steps: (1) determine if calibration is needed, and (2) calibrate if needed. In the first step, known input signals (e.g., 0, 25, 50, 75, and 100% of span) are applied to the instrument, and its output is recorded on a data sheet. The data thus generated is referred to as the as-found calibration data (see Table 1.8e). If the asfound data show that the instrument’s calibration is still acceptable, no calibration is needed. Otherwise, the instrument is
to make the electrical output proportional to the applied pressure. For example, a pressure sensor may be calibrated to produce an output in the range of 4 to 20 mA for pressure inputs covering the whole span of the sensor (0 to 100%). For most pressure sensors, with no pressure applied, the transmitter output is adjusted to produce a 4-mA signal. Next,
http://abzardaghigh.ir/duh/doc_download/141-instrument-calibration.html
|
پی ال سی زیمنس به زب...
جهت دانلود همه مقالات سایت ابتدا لاگی...
پی ال سی زیمنس به زب...
re: plc زيمنس بزبان فارسي - سرکا...
انواع روشهای اندازه ...
یک سوال - با سلام : مخواستم بدونم فر...
جزوه آموزشي plc وزار...
جزوه آموزشي plc وزارت كار - درود ب...
پی ال سی زیمنس به زب...
plc زيمنس بزبان فارسي - درود بر ...