|
Response Time and Drift Testing
Off-line calibration of the zero and span of measurement was the topic of the previous section. In this section, the on-line methods of response time determination and calibration verification will be described for sensors that have already been installed in operating processes. As in Section 1.8, the discussion here will also focus on temperature and pressure sensors. FUNDAMENTALS OF RESPONSE TIME TESTING The response time of an instrument is measured by applying a dynamic input to it and recording the resulting output. The recording is then analyzed to measure the response time of the instrument. The type of analysis is a function of both the type of instrument under test and on the type of dynamic input applied, which can be a step, a ramp, a sine wave, or even just random noise. The terminology used in connection with time response to a step change was defined in Figure 1.3z. The time constant ( T ) of a first-order system was defined as the time required for the output to complete 63.2% of the total rise (or decay) resulting from a step change in the input. Figures 1.9a and 1.9b show the responses of instruments to both step changes and ramps in their inputs and identify the time constant ( T ) and response times ( τ ) of these instruments. As shown in Figure 1.9a, the time constant of an instrument that responds as a first-order system equals its response time and it is determined by measuring, after a step change in the input, the time it takes for the output to reach 63.2% of its final value. The response of a first-order system is mathematically described by a first-order differential equation,
Although most instruments are not first-order systems, their response time is often determined as if they were, and as if their response time were synonymous with their time constant. However, if the system is of higher than first order, there is a time constant for each first-order component in the system. In spite of this, in the field, the definition of the firstorder time constant is often also used in connection with higher-order systems. The ramp response time is the time interval by which the output lags the input when both are changing at a constant rate. For a ramp input, the response time ( τ ) is defined as the delay shown in Figure 1.9b. This is also referred to as ramp time delay and can be measured after the initial transient, when the output response has become parallel with the input ramp signal. For a first-order system, the ramp time delay, response time, and time constant are synonymous. The ramp time delay can be mathematically described as
where C is the ramp rate of the input signal. The derivations of Equations 1.9(1) through 1.9(3) and the topic of Laplace transformation is covered in the second volume of the Instrument Engineers’ Handbook and also in Reference 1. LABORATORY TESTING The response time of temperature sensors is measured by using a step input, whereas the response time of pressure sensors is usually detected by using ramp input signals. This is because obtaining a step change in temperature is easier and more repeatable than obtaining a step change in pressure. Ramp inputs are also preferred for the testing of pressure sensors, because a step input can cause oscillation of the pressure transmitter output, which may complicate the measurement. Testing of Temperature Sensors Figure 1.9c illustrates the equipment used in determining the response time of a temperature sensor. This experiment is called the plunge test. At the beginning of the test, the sensor is held by a hydraulic plunger, and its output is connected to a recorder. The heated sensor is then plunged into a tank of water at near-ambient temperature. This step change in temperature determines the type of transient in its output, as was illustrated in Figure 1.9a. To identify the response time of the temperature sensor, the time corresponding to 63.2% of the full response is measured.
Because the response time of a temperature sensor is a function of the type, flow rate, and temperature of the media in which the test is performed, the American Society for Testing and Material (ASTM) has developed Standard E644 (Reference 2), which specifies a standard plunge test. This document specifies that a plunge test should be performed in water that is at near room temperature and is flowing at a velocity of 3 ft/sec (1 m/sec). A plunge test can therefore be performed by heating the sensor and then plunging it into a rotating tank that contains water at room temperature. By controlling the speed and the radial position of the sensor, the desired water velocity can be obtained for the plunge test. There can be other ways for performing the plunge test. For example, the sensor can be at room temperature and plunged into warm water. Although the actual temperatures have an effect on response time, this effect is usually small; therefore, the response time is not significantly different if the water is at a few degrees above or below room temperature. Testing of Pressure Sensors The response time of pressure sensors is usually determined by using hydraulic ramp generators, which produce the ramp test input signals. A photograph of a hydraulic ramp generator is provided in Figure 1.9d. This equipment consists of two pressure bottles, one bottle filled with gas or air and the other
http://abzardaghigh.ir/duh/doc_download/142-response-time-and-drift-testing.html
|
پی ال سی زیمنس به زب...
جهت دانلود همه مقالات سایت ابتدا لاگی...
پی ال سی زیمنس به زب...
re: plc زيمنس بزبان فارسي - سرکا...
انواع روشهای اندازه ...
یک سوال - با سلام : مخواستم بدونم فر...
جزوه آموزشي plc وزار...
جزوه آموزشي plc وزارت كار - درود ب...
پی ال سی زیمنس به زب...
plc زيمنس بزبان فارسي - درود بر ...